KYOTO ELECTRONICS MANUFACTURING CO., LTD.

PRODUCT

Karl Fischer Moisture Titrator (Volumetric titration) [MKV-710D]

MKV-710D-en

*The photo includes 'Additional Burette KF(10mL)'(option). A single burette is equipped as standard.

MKV-710D is a Karl Fischer Moisture Titrator equipped with data integrity support functions without a PC. Simultaneous measurement can be performed with up to two units and the large color LCD touch panel can be linked wirelessly. Your MKV-710M or MKV-710S can be upgraded to MKV-710D.

Features

  • Dl support function
    Access Restrictions :Operator authentication by operator ID and password
    Rights Administration :Each operator is assigned to a specific right group
    Audit Trail :Operation histories are automatically recorded
    Electronic Approval / Electronic Signature :Electronic approval and signature based on the workflow
    Backup / Restoration :Automatic backup with the predetermined period
    Archive / Archive View :Data storage. Archive data can be viewed on the main control unit
    Data Capacity :The remaining capacity can be checked any time via an icon on the main screen
    Security Settings :Password length and lockout settings can be set
    LIMS Connection :Data output to LIMS systems
  • Large color TFT-LCD with a touch panel (8.4-inch)
  • The main control unit and a titrator are separately located by wireless (Bluetooth(R)).
  • Measurement results can be saved in PDF files.
  • The Water content can be displayed in real time.
  • Burettes can be selected from 1-unit and 2-unit types
  • Not necessary to change the electrode sensitivity and endpoint potential

Specifications

Type Karl Fischer Moisture Titrator
Model MKV-710D
Standard configuration MCU-710D+MKV-710+Automatic Solvent Change unit
Measuring method Karl Fischer Volumetric titration
Measuring range 1) Water content : 0.1 to 500mgH2O (depends on KF reagent factor)
2) Concentration : 1ppm to 100%H2O
Burette precision 10mL burette +/- 0.015mL ; reproducibility +/- 0.005mL
Endpoint detection By polarized potential level detected with a twin platinum electrode
EP sense method Detection of potential level maintained during preset end time.
End time range : 1 to 99s
Titration form Normal titration / Back titration (Option additional burette required)
Required solvent 30 to 100mL (in S-type titration vessel)
Key operation Touch panel
Displays 1) 8.4-inch color LCD 800 x 600 dots
2) English / Japanese / Mandarin Chinese
3) Simultaneous 2-channel display
Calculation Concentration of water content , statistics data processing (mean , SD and RSD) and automatic averaging of blank value and factor value
Data storage Method: 120 (per channel)
Measurement results: 5,000 samples (per channel)
Operation history: 100,000 (Total for 2 channels)
DI support function Access Restrictions: Operator authentication by operator ID and password
Rights Administration: Each operator is assigned to a specific right group
Audit Trail: Operation history is automatically recorded
Electronic Approval / Electronic Signature: Electronic approval and signature based on the workflow
Backup / Restoration: Automatic backup with the predetermined period
Archive / Archive View: Data storage. Archived data can be viewed on the main control unit
Data Capacity: The remaining capacity can be checked any time via an icon on the main screen
Security Setup: Password length and lockout settings can be set
LIMS Connection: Data output to LIMS systems
External I/O RS-232C port x4 :
for Dot matrix printer , Electronic balance , Data Capture Software (SOFT-CAP) , Evaporator
USB ×1 :
for USB flash drive , A4 printer , Thermal printer , Keyboard , Barcode reader , Foot switch , USB HUB
LAN ×1 : for Personal computer (PC)
Extensibility Measuring instrument :
Karl Fischer Moisture Titrator (MKV-710/MKC-710)
One of those instruments can be added
Automatic piston burette : Can control max 2 burette drives (Including two built-in burette drives)
Evaporator ADP-611
Ambient condition
1) Temperature : 5 to 35°C
2) Humidity : 85%RH or below (no condensation)
Power source AC100 to 240V +/- 10% 50/60 Hz
Power consumption
Main unit : Approx. 30W
Dimensions
Touch panel controller : 225(W) x 190(D) x 42(H) mm
Titration unit : 141(W) x 292(D) x 367(H) mm (not incl. tubing)
Stirrer : 107(W) x 206(D) x 322(H) mm (not incl. Solvent Change unit)
Solvent Change Unit : 240(W) x 140(D) x 400(H) mm (not incl. tubing)
Weight
Touch panel controller : Approx.1.5kg
Titration unit : Approx.4.0kg
Stirrer : Approx.2.0kg
Solvent Change Unit : Aprox.0.6kg
Conformity standard CE marking EMC : EN61326-1 LVD : EN61010-1 RE Directive
Burette unit EBU FCC Part15 SubpartC FCC ID : 2ABSVEBU01